Cardiovascular Research Advance Access originally published online on October 12, 2007
Cardiovascular Research 2008 77(1):81-88; doi:10.1093/cvr/cvm036
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Transmural temporospatial left ventricular activation during pacing from different sites: potential implications for optimal pacing
Department of Cardiac Sciences and the Libin Cardiovascular Institute, University of Calgary, Health Sciences Centre, 3330 Hospital Drive NW, Calgary, Alberta, Canada T2N 4N1
* Corresponding author. Tel: +1 403 210 8838; fax: +1 403 283 6151. E-mail address: kkavanag{at}ucalgary.ca
Received 24 April 2007; revised 6 October 2007; accepted 10 October 2007
Time for primary review: 30 days
| Abstract |
|---|
|
|
|---|
Aims: Previous studies showed that right ventricular (RV) endocardial pacing can be deleterious even in individuals with initially normal left ventricular (LV) function. The mechanism(s) by which RV endocardial pacing may cause LV dysfunction is unknown. This study compares the temporospatial LV transmyocardial activation profiles during sinus rhythm with normal His/Purkinje conduction vs. currently utilized and proposed cardiac pacing sites.
Methods and results: Mongrel dogs were instrumented with transmural electrodes that tracked transmyocardial activation sequences at five sites in the LV. Pacing/recording catheters were positioned in the RV apex and on the RV and LV sides of the ventricular septum. An epicardial pacing electrode was also sewn to the mid-lateral LV epicardium. Electrograms were recorded during sinus rhythm and pacing from the RV endocardium, LV septum, LV epicardium and during biventricular pacing. Compared to normal sinus/His/Purkinje rhythm (NSR), RV endocardial pacing significantly (P < 0.05) prolonged transmural activation (NSR endocardium 6.1 ± 1 ms vs. RV endocardium 23.0 ± 2.6 ms). The highly ordered temporospatial pattern of transmural activation during sinus rhythm was replaced with dispersion and intermingling of endo-, mid-, and epicardial activation. LV epicardial and biventricular pacing did not correct these abnormalities. Only LV septal pacing achieved the transmural and transseptal activation sequences similar to sinus rhythm.
Conclusion: Clinically utilized pacing modalities, including biventricular pacing, cause abnormal transmyocardial activation. LV septal pacing results in transmyocardial activation patterns that closely resemble those seen in sinus rhythm.
KEYWORDS Ventricular pacing; Transmural activation; Septal activation; Biventricular pacing
| 1. Introduction |
|---|
|
|
|---|
More than 200 000 pacemakers are implanted annually in North America1 with the most frequent indication being sinus node dysfunction (58.8%).2 Dual-chamber pacing was developed to restore and maintain atrioventricular (AV) synchrony and is considered a physiologic mode.3–6 However, trial results relative to heart failure have been inconsistent suggesting either modest or no benefit in limiting progression to heart failure.4–7 Analyses of randomized trials suggested that right ventricular (RV) pacing, with or without maintenance of AV synchrony, can lead to higher rates of new or progressive heart failure and increased mortality.6,8 This was recently verified in the MOST trial9–11 which was the largest (2010 patients) randomized trial to compare single-chamber vs. dual-chamber pacing (the RV was paced in both) in patients with sinus node dysfunction.4 Heart failure occurred least frequently in patients randomized to dual-chamber pacing but with a very low cumulative % of RV pacing.5 Despite maintaining AV synchrony, dual chamber pacing with >40% ventricular pacing was associated with a 2.6-fold increase in risk of hospitalization for heart failure.5,9,10 Similarly, during single-chamber RV pacing with a high cumulative per cent pacing, the probability of heart failure increased by nearly 40-fold.5 Thus, a new (untested) strategy has been recommended for physiologic ventricular pacing—if ventricular or AV conduction is abnormal, RV septal, left ventricular (LV) epicardial, or biventricular pacing should be considered instead of RV apical pacing.3
This canine study systematically compares LV temporospatial transmyocardial activation profiles in normal sinus/His/Purkinje rhythm (NSR) with both clinically utilized and proposed pacing sites with a view to identifying the site(s) which result in the most physiological transmyocardial activation.
| 2. Methods |
|---|
|
|
|---|
2.1 Animal preparation
Five mongrel dogs weighing 20–25 kg were premedicated with 25 mg/kg sodium thiopental. Anaesthesia was maintained with intravenous fentanyl citrate (0.04 mg/mL; given to effect), followed by an infusion of
4 mg/h which was adjusted as necessary to maintain a surgical plane of anaesthesia. The animals were intubated and ventilated with a constant volume ventilator (Harvard apparatus, Natick, MA, USA) with 50% O2/50% N2O. Through a median sternotomy, the pericardium was opened with a base-to-apex incision. The tip of a quadrapolar-pacing catheter was positioned in the RV apex via the right external jugular vein under fluoroscopic guidance. Medtronic 5076 52 cm active fixation leads (6 F) were inserted through the LV and RV free walls and screwed into the LV and RV sides of the septum, respectively, using fluoroscopic guidance. Five short bevelled 20 gauge needles, each containing 3 Ag/AgCl electrodes with a 4 mm inter-electrode distance (2 electrodes in the shaft and the third in the phenolic base) were inserted into the LV (Figure 1) [anterior base (AB), the posterior base (PB), the lateral free wall (LFW) halfway between apex and base, the mid-anterior free wall (AFW), and the LV apex]. An Ag/AgCl epicardial pacing electrode was sutured to the middle of the lateral LV free wall. The pericardium was then closed, and chest wall was opposed. A 6 F catheter was inserted into the femoral artery for continuous arterial pressure monitoring and to obtain samples for blood gas analysis. A three-lead electrocardiogram was recorded throughout procedure.
|
All studies were approved by the institutional Animal Welfare Committee at the University of Calgary and carried out in accordance with the Canadian Council on animal care guidelines as well as the Guide for the Care and Use of Laboratory Animals published by the US National Institutes of Health (NIH Publication No. 85-23, revised 1996).
2.2 Experimental protocol
Electrograms were recorded during NSR, RV endocardial (septal and apical), LV epicardial, LV septal, and biventricular pacing (RV septal+LV epicardial pacing). Each site was paced at twice diastolic threshold for 2 min at a rate just above the sinus rate so that there was 1:1 capture. Pacing sequences were randomized to minimize the potential impact of time-dependent changes. There was a minimum interval of 5 min between pacing from each site. Electrograms recorded with the distal plunge electrodes were considered to be recorded from the endocardium. Electrograms recorded with the mid-plunge electrodes were considered to be recorded from the mid-myocardium and those recorded with the proximal or surface plunge electrode were categorized as epicardial in origin.
2.3 Data acquisition
Unipolar signals from each recording electrode were referenced to the chest wall using a 4 mm sintered disk electrode and DC coupled, amplified, filtered (0.2–5000 Hz), simultaneously sampled at 16 kHz, and analogue-to-digitally converted with 12 bits of resolution.12 The input impedance of the amplifiers was 1012
. Unipolar electrogram signals were stored together with ECG signals. Five-point, second-order data-fit was used to calculate all derivatives.13 Location of peak negative dV/dt (maximum change in voltage over time) for unipolar electrograms was automatically determined and was used to indicate time of activation.12,13 Myocardial regional activation time was defined as the mean of the electrogram activation times recorded by the plunge electrodes in the individual regions (e.g. mean of endocardial recordings, etc). The total duration of myocardial activation was defined as the length of time between the earliest and latest myocardial activation. The envelope of activation is defined as the total duration of myocardial activation as well as the spatial sequence and duration of activation of endocardium, mid-myocardium, and epicardium. Regional myocardial dispersion of activation was defined as the time difference between earliest and latest activation in each myocardial region; that is endo- is separately analysed from mid- and epi- data.
2.4 Statistical analysis
Data are expressed as mean±SE unless otherwise indicated. Mean values were compared between NSR and the different pacing sites using ANOVA followed by the Dunnett correction factor. Differences with P < 0.05 were considered statistically significant.
| 3. Results |
|---|
|
|
|---|
3.1 Envelope of activation
Figure 2A–E shows typical transmural electrograms recorded during NSR and during pacing at each site. Left column shows transmural electrograms referenced to the surface ECG. The middle and right columns show expanded views of the transmural electrograms during depolarization and repolarization, respectively. During NSR (Figure 2A, middle column), activation of endocardium (red) was rapid; activation of the mid-myocardium (blue) occurred soon after and was complete prior to epicardial activation (black).
|
LV septal pacing (Figure 2B, middle column) produced an envelope of activation that was very similar to that in NSR. In addition to similar endo-, mid-, epicardial, and total activation durations, the transmural spatial activation sequence was also similar. This was not true with biventricular (Figure 2C, middle column), LV epicardial (Figure 2D, middle column), and RV endocardial (Figure 2E, middle column) pacing where the total duration of activation and the activation duration of individual regions were prolonged compared with NSR and LV septal pacing. In addition, the organized sequence of activation observed during NSR (Figure 2B, middle column) was replaced by intermingling of endo-, mid-, and epicardial activation (Figures 2C–E, middle column). RV endocardial pacing (Figure 2E, middle column) caused the most prolonged regional activation durations as well as the most disorganized sequence of activation.
The right column in Figure 2 shows the envelopes of repolarization associated with the envelopes of activation in the middle column. The signal-to-noise ratio is much smaller for repolarization than depolarization. As expected, repolarization varies with the different pacing modes. Figure 3 shows representative samples of repolarization during NSR and the various pacing modes. The signal-to-noise ratio is very poor. Even during NSR, there is variability in repolarization (Figure 3A) and, as anticipated, the variability is even more pronounced during pacing (Figure 3B–E).
|
3.2 Transmural activation sequence
Transmural activation sequence is more formally presented as a radial plot for NSR and for each pacing site (Figure 4). Each arm of the plot represents an anatomical recording site. During NSR (Figure 4A), the LV septum (red symbol) was activated first followed rapidly by activation of the remaining LV endocardium (red symbols) and sequential activation of mid-myocardium (blue symbols) and epicardium (black symbols). During NSR, the RV septum was activated later than all the LV endocardial sites. With LV septal pacing (Figure 4B), transmural activation was very similar to that during NSR. With biventricular pacing (Figure 4C), LV endocardial activation was markedly prolonged compared to that during NSR. There was intermingling of endo-, mid-, and epicardial activation with some epicardial areas predictably activating before some endocardial areas. LV endocardial activation was also prolonged during LV epicardial pacing (Figure 4D) compared to that during NSR. There was also intermingling of endo-, mid-, and epicardial activation. Pacing from RV endocardial sites caused prolonged and disordered (intermingling of endocardial, mid-myocardial, and epicardial activation) transmural LV activation. LV septal activation was markedly delayed compared to that in NSR.
|
3.3 Mean transmyocardial activation times
Figure 5 shows the mean transmural regional activation times (± SE) for NSR and for each pacing site. Endocardial activation during NSR was significantly more rapid (6.1 ± 1.0 ms) than during pacing at the other sites (RV endo, 23.0 ± 2.6 ms; LV epi, 15.3 ± 2.2; BIV, 14.5 ± 1.9 ms; P < 0.05) with the exception of LV septal pacing (11.7 ± 2.6 ms; NS compared to NSR). Importantly, the pattern of transmural activation during NSR and LV septal pacing was similar. The activation wave sequentially and progressively moved from endo- to mid- to epi- sites in NSR and during LV septal pacing. Mean endocardial activation times were 6.1 ± 1.0 ms, for NSR; and 11.7 ± 2.6 ms, for LV septal pacing. Mean mid-myocardial activation times were 10.7 ± 0.8 ms for NSR and 15.9 ± 2.8 ms for LV septal pacing (P < 0.05 compared to endocardium) while mean epicardial activation times were 21.6 ± 0.8 ms for NSR and 24.9 ± 3.1 ms for LV septal pacing (P < 0.05 compared to mid-myocardium). Because of intermingling of activation in the endo-, mid-, and epicardial regions, this sequential and progressive transmyocardial activation relationship did not exist with pacing at the other sites.
|
During NSR, no time-dependent changes were observed in QRS amplitude in Lead II (0.57 ± 0.20 mV at the start of the protocol vs. 0.46 ± 0.13 mV at the end, P=ns) or duration (49.8 ± 6.3 ms at the start vs. 49.1 ± 5.5 ms at the end, P=ns). There was no significant difference in arterial pressure during NSR and the different modes of pacing.
3.4 Dispersion of activation
Figure 6 shows the mean dispersion of activation (± SE) for endo-, mid-, and epicardial regions during NSR, LV septal, biventricular, and RV endocardial pacing. The greatest dispersion of activation in all three myocardial regions (endo-, mid-, epi-) was observed during RV endocardial pacing. There was no significant difference in dispersion of activation between RV endocardial pacing and biventricular pacing. The least dispersion of activation occurred during NSR. Dispersion of activation in all three myocardial regions (endo-, mid- and epi-) was significantly greater during RV endocardial pacing (30.1 ± 5.4 ms for endo; 39.4 ± 3.9 ms for mid; and 45.5 ± 2.2 ms for epi) compared to either NSR (5.9 ± 0.8 ms for endo; 6.6 ± 0.8 ms for mid; and 7.0 ± 0.4 ms for epi) or LV septal pacing (13.0 ± 5.3 ms for endo; 16.3 ± 5.4 ms for mid; and 17.7 ± 5.5 ms for epi).
|
3.5 Septal activation
Figure 7 shows RV and LV septal activation times during NSR and during pacing at the other sites. The LV and RV septal activation relations and times were very similar during NSR and LV septal pacing. During RV endocardial and biventricular pacing, the septal activation sequence was reversed with the RV septum being activated initially and LV septal activation being significantly delayed compared to during NSR. In fact, during RV endocardial pacing, LV septal activation time was similar to that of the LV free wall endocardium (Figure 4E). The LV septum was the last endocardial site to be activated during biventricular pacing (Figure 4C).
|
| 4. Discussion |
|---|
|
|
|---|
In the present study, we have shown that the timing, sequence, and duration of myocardial activation vary substantially with pacing sites. This work adds to that of Medina-Ravell et al.14 with respect to LV epicardial and biventricular pacing. During NSR, activation is rapid and uniform with the LV endocardium being activated first, followed by the mid-myocardium and then, the epicardium. However, activation duration and sequence are substantially altered during pacing at all other sites except the LV septum, which resulted in a transmyocardial activation duration and sequence similar to those observed during NSR. RV endocardial pacing caused marked prolongation and a disordered sequence of endo-, mid-, and epicardial activation. Biventricular pacing did not correct these abnormalities. Additionally, septal activation was reversed during both RV endocardial and biventricular pacing compared to NSR.
4.1 Envelope of activation
The importance of the temporospatial activation pattern for optimal myocardial performance was recognized by Wiggers15 and has been referred to as the idioventricular kick.16 Assessment of LV activation should include total duration as well as duration of endocardial, mid-myocardial, and epicardial activation and the sequence of transmural activation. The most obvious difference between activation through the normal His-Purkinje system and RV endocardial pacing is the prolonged duration of activation. Ventricular activation via the Purkinje network normally occurs within approximately 40 ms, whereas with RV endocardial pacing, the altered activation sequence creates delays of up to 100 ms.17 During RV endocardial, LV epicardial, and biventricular pacing, the spatial activation is substantially different than during NSR. The progression of activation from endocardium to mid-myocardium to epicardium is replaced by an intermingling of activation in all three regions.
The consequences of long-term RV endocardial pacing are potentially harmful. Myocardium which is activated early can stretch not-yet-depolarized muscle, which in turn, when activated, can stretch repolarized earlier activated segments.16–18 The resulting contraction is thus mechanically and metabolically inefficient.19 Reduced myocardial blood flow and increased wall thickness have been demonstrated in early-activated regions.17,18 This can result in LV remodelling with asymmetric hypertrophy, mitral regurgitation, decreased ejection fraction, and increased left atrial size.3,16,17 These changes appear to be reflected in clinical outcomes—despite maintaining AV synchrony, patients with the greatest amount of pacing are hospitalized more often for heart failure than those who are paced less.5
Not surprisingly, during this acute study in animals with normal hearts, the brief periods of pacing did not produce any significant changes in aortic pressure despite the changes in the transmural envelope of activation. LV epicardial pacing showed intermingling of endo-, mid-, and epicardial activation rather than the normal activation sequence seen during NSR. Endocardial activation was also prolonged and there was marked epicardial dispersion of activation. Resynchronization therapy, which combines RV endocardial and LV epicardial pacing to optimize mechanical synchrony, is recommended in patients with moderate to severe LV dysfunction who require pacing or have poorly controlled heart failure.20,21 During biventricular pacing, there was also intermingling of endo-, mid-, and epicardial activation; endocardial activation was slower than during NSR, though faster than during RV endocardial pacing. Clinically, in patients with abnormal LV dysfunction, biventricular pacing is ineffective in
30% of patients.22 It is possible that the sequence of transmyocardial activation plays a role in limiting the effectiveness of this therapy.
4.2 Septal activation
During normal conduction, the left mid-septal endocardium is activated first23,24 in keeping with our findings. Purkinje fibres do not penetrate the septum and conduction through the remainder of the septum is syncytial.24 Pacing from a point source results in the contralateral ventricle not being activated normally and the impulse may enter the contralateral ventricle through multiple pathways.25 During RV endocardial pacing, the septal activation sequence was reversed with left septal activation being markedly delayed compared to during NSR. The earliest LV activation commonly occurred close to the septum, either anterior or posterior, before activation of the mid-LV septum. LV epicardial activation resulted in delayed RV and LV septal activation. Biventricular pacing resulted in a septal activation pattern similar to RV endocardial pacing. Thus, these results suggest that optimal septal activation is not achieved with presently used or proposed clinical pacing sites.
4.3 Left ventricular septal pacing
LV septal pacing achieved a temporospatial activation envelope that closely resembled that during NSR, whereas pacing at all other study sites, including biventricular pacing, did not. No changes in aortic pressure were demonstrated during the short episodes of pacing in the various pacing modes. Peschar et al.,26 who also paced the LV septum in dogs demonstrated no deterioration in stroke work, end-systolic, and end-diastolic volumes compared with NSR but were unable to explain the favourable haemodynamics as the QRS duration was still prolonged compared with NSR.26 Effective LV septal pacing has been achieved experimentally,27 which suggests that such a strategy may be clinically feasible. Since pacing from the LV septal site produces an envelope of activation that most closely resembles normal sinus rhythm, further work to determine if LV septal pacing can prevent or minimize the impact of pacing on the development or progression of heart failure appears justified.
4.4 Limitations
Relatively few recording electrodes were used in this study; thus there may be earlier and later myocardial activation sites than those recorded. However, this study did include transmural and septal recordings during NSR and pacing from presently utilized and proposed sites. Our results during NSR were similar to those previously reported23–25 and demonstrated the disarray of temporospatial patterns of activation associated with pacing. It has previously been shown that even large numbers of plunge electrodes do not alter activation sequence.28
This was an acute study and episodes of pacing in all modalities were brief, thus the potential long-term consequences of structural remodelling and its haemodynamic effects, though critically important, could not be addressed by this study.
| Funding |
|---|
|
|
|---|
Supported by a grant from the Heart and Stroke Foundation of Canada/Alberta.
| Acknowledgements |
|---|
The authors wish to thank Cheryl Meek and Keith Robinson for their excellent technical assistance.
Conflict of interest: none declared.
| References |
|---|
|
|
|---|
- Lamas GA, Lee K, Sweeney M, Leon A, Yee R, Ellenbogen K, et al. The mode selection trial (MOST) in sinus node dysfunction: design, rationale, and baseline characteristics of the first 1000 patients. Am Heart J (2000) 140:541–551.[CrossRef][Web of Science][Medline]
- Birnie D, Williams K, Guo A, Mielniczuk L, Davis D, Lemery R, et al. Reasons for escalating pacemaker implants. Am J Cardiol (2006) 98:93–97.[CrossRef][Web of Science][Medline]
- Sweeney MO, Prinzen FW. A new paradigm for physiologic ventricular pacing. J Am Coll Cardiol (2006) 47:282–288.
[Abstract/Free Full Text] - Lamas GA, Lee KL, Sweeney MO, Silverman R, Leon A, Yee R, et al. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med (2002) 346:1854–1862.
[Abstract/Free Full Text] - Sweeney MO, Hellkamp AS, Ellenbogen KA, Greenspon AJ, Freedman RA, Lee KL, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation (2003) 107:2932–2937.
[Abstract/Free Full Text] - Steinberg JS, Fischer A, Wang P, Schuger C, Daubert J, McNitt S, et al. The clinical implications of cumulative right ventricular pacing in the multicenter automatic defibrillator trial II. J Cardiovasc Electrophysiol (2005) 16:359–365.[Web of Science][Medline]
- Nielsen JC, Kristensen L, Andersen HR, Mortensen PT, Pedersen OL, Pedersen AK. A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. J Am Coll Cardiol (2003) 42:614–623.
[Abstract/Free Full Text] - Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA (2002) 288:3115–3123.
[Abstract/Free Full Text] - Sweeney MO, Hellkamp AS. Heart failure during cardiac pacing. Circulation (2006) 113:2082–2088.
[Abstract/Free Full Text] - Shukla HH, Hellkamp AS, James EA, Flaker GC, Lee KL, Sweeney MO, et al. Heart failure hospitalization is more common in pacemaker patients with sinus node dysfunction and a prolonged paced QRS duration. Heart Rhythm (2005) 2:245–251.[CrossRef][Web of Science][Medline]
- Sweeney MO, Hellkamp AS, Lee KL, Lamas GA. Association of prolonged QRS duration with death in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation (2005) 111:2418–2423.
[Abstract/Free Full Text] - Kavanagh KM, Guerrero PA, Jugdutt BI, Witkowski FX, Saffitz JE. Electrophysiologic properties and ventricular fibrillation in normal and myopathic hearts. Can J Physiol Pharmacol (1999) 77:510–519.[CrossRef][Web of Science][Medline]
- Witkowski FX, Penkoske PA. A new fabrication technique for directly coupled transmural cardiac electrodes. Am J Physiol (1988) 254:H804–H810.[Web of Science][Medline]
- Medina-Ravell VA, Lankipalli RS, Yan GX, Antzelevitch C, Medina-Malpica NA, Medina-Malpica OA, et al. Effect of epicardial or biventricular pacing to prolong QT interval and increase transmural dispersion of repolarization: does resynchronization therapy pose a risk for patients predisposed to long QT or torsade de pointes? Circulation (2003) 107:740–746.
[Abstract/Free Full Text] - Wiggers CJ. The muscular reactions of the mammalian ventricles to artificial surface stimuli. Am J Physiol (1925) 73:346–378.
[Free Full Text] - Stevenson WG, Sweeney MO. Single site left ventricular pacing for cardiac resynchronization. Circulation (2004) 109:1694–1696.
[Free Full Text] - Faris OP, Evans FJ, Dick AJ, Raman VK, Ennis DB, Kass DA, et al. Endocardial versus epicardial electrical synchrony during LV free-wall pacing. Am J Physiol Heart Circ Physiol (2003) 285:H1864–H1870.
[Abstract/Free Full Text] - Prinzen FW, Hunter WC, Wyman BT, McVeigh ER. Mapping of regional myocardial strain and work during ventricular pacing: experimental study using magnetic resonance imaging tagging. J Am Coll Cardiol (1999) 33:1735–1742.
[Abstract/Free Full Text] - Baller D, Wolpers HG, Zipfel J, Bretschneider HJ, Hellige G. Comparison of the effects of right atrial, right ventricular apex and atrioventricular sequential pacing on myocardial oxygen consumption and cardiac efficiency: a laboratory investigation. Pacing Clin Electrophysiol (1988) 11:394–403.[CrossRef][Medline]
- Abraham WT, Young JB, Leon AR, Adler S, Bank AJ, Hall SA, et al. Effects of cardiac resynchronization on disease progression in patients with left ventricular systolic dysfunction, an indication for an implantable cardioverter-defibrillator, and mildly symptomatic chronic heart failure. Circulation (2004) 110:2864–2868.
[Abstract/Free Full Text] - Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L, et al. Longer-term effects of cardiac resynchronization therapy on mortality in heart failure [the CArdiac REsynchronization-Heart Failure (CARE-HF) trial extension phase]. Eur Heart J (2006) 27:1928–1932.
[Abstract/Free Full Text] - Daubert JC, Leclercq C, Donal E, Mabo P. Cardiac resynchronisation therapy in heart failure: current status. Heart Fail Rev (2006) 11:147–154.[CrossRef][Web of Science][Medline]
- Durrer D, van Dam RT, Freud GE, Janse MJ, Meijler FL, Arzbaecher RC. Total excitation of the isolated human heart. Circulation (1970) 41:899–912.
[Abstract/Free Full Text] - Scher AM, Young AC, Malmgren AL, Erickson RV. Activation of the interventricular septum. Circ Res (1955) 3:56–64.
[Abstract/Free Full Text] - Lister JW, Klotz DH, Jomain SL, Stuckey JH, Hoffman BF. Effect of pacemaker site on cardiac output and ventricular activation in dogs with complete heart block. Am J Cardiol (1964) 14:494–503.[CrossRef][Web of Science][Medline]
- Peschar M, de Swart H, Michels KJ, Reneman RS, Prinzen FW. Left ventricular septal and apex pacing for optimal pump function in canine hearts. J Am Coll Cardiol (2003) 41:1218–1226.
[Abstract/Free Full Text] - Grosfeld MJ, Res JC, Vos DH, de Boer TJ, Bos HJ. Testing a new mechanism for left interventricular septal pacing: the transseptal route; a feasibility and safety study. Europace (2002) 4:439–444.
[Abstract/Free Full Text] - Kovoor P, Campbell C, Wallace E, Byth K, Dewsnap B, Eipper V, et al. Effects of simultaneous insertion of 66 plunge needle electrodes on myocardial activation, function, and structure. Pacing Clin Electrophysiol (2003) 26:1979–1985.[CrossRef][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||







P < 0.01.