The volume-dependency of parallel conductance throughout the cardiac cycle and its consequence for volume estimation of the left ventricle in patients
aDepartment of Physiology, Cardiovascular Research Institute Maastricht, Maastricht University, P.O. Box 616, 6200 MD Maastricht, The Netherlands
bDepartment of Cardiac Surgery, San Raffaele Hospital, Milan, Italy
cDepartment of Cardiology, Leiden University Medical Center, Leiden, The Netherlands
dDepartment of Intensive Care, Leiden University Medical Center, P.O. Box 9600, 2300 RC Leiden, The Netherlands
l.kornet{at}fys.unimaas.nl
* Corresponding author. Tel.: +31-43-388-1209; fax: +31-43-388-4166
Received 19 February 2001; accepted 13 April 2001
| Abstract |
|---|
|
|
|---|
Objective: To study the hypothesis that the electrical conductance of tissues and fluids (parallel conductance (Gp)) around the ventricle depends on left ventricular volume throughout the cardiac cycle. Methods: We extended a recently developed method to determine Gp throughout the cardiac cycle. First, we compared the estimates of parallel conductances obtained with the new method (Gap) with those of the conventional one (Glp), both averaged over the cardiac cycles. Secondly, Gap was determined throughout the cardiac cycle and its volume dependency was assessed. Thirdly, the factor
was calculated as the ratio between stroke volume, obtained by the conductance method using Glp, and that obtained by a thermodilution method. Because the non-homogeneous field was indicated to be the reason for the dependency of Gp on left ventricular volume as well as for the need for
, we tested whether the hypothesis implies that a correction with
is not needed if Gp is determined throughout the cardiac cycle. Results: We found a negative linear relation between Gp and left ventricular volume. This relation appeared to be reproducible within each patient. Furthermore, we found that
deviates from 1 primarily due to the dependency of Gp on left ventricular volume. Conclusion: To obtain stroke volume or to determine absolute left ventricular volume continuously within a cardiac cycle, Gp should be determined throughout each cardiac cycle and if a constant Gp throughout the cardiac cycle is used a correction with the factor
should be made to correct for a possible influence of electrical field heterogeneity.
KEYWORDS Cardiomyopathy; Cardiovascular surgery; Contractile function; Heart failure; Ventricular function
| 1 Introduction |
|---|
|
|
|---|
We expect that in the near future the assessment of pressure–volume relationships in both ventricles will become more and more important in the determination of cardiac function in a clinical setting. Therefore, besides pressure, absolute ventricular volume should be determined continuously. The electrical conductance catheter has been used to measure the volume of a cardiac chamber continuously by positioning a catheter in the right or left ventricle and relating the electrical conductance of blood in the ventricle to its volume [1–20]. Besides blood, the tissues surrounding the ventricle contribute to the measured conductance and cause an offset in the conductance signal, i.e. parallel conductance (Gp). Conventionally parallel conductance is determined by plotting values at end-systole (Gsys) versus those at the preceding end-diastole (Gdias) during a beat-to-beat increase in conductivity of blood after the injection of hypertonic saline. The linear regression line through Gsys–Gdias values is extrapolated to the identity line to estimate parallel conductance (Glp) [2]. For this estimation it is assumed that Glp is constant within a cardiac cycle. Theoretical studies, however, predict a dependence of parallel conductance on volume [21–23]. If the ventricle becomes larger, the density of electric current at the myocardial boundary and therefore parallel conductance will be smaller due to the non-homogeneous electrical field. If a constant Glp during a cardiac cycle is assumed as with the conventional conductance method, stroke volume, estimated with the conductance catheter, was different from stroke volume obtained by an independent method [1,2,7]. A correction factor
, being the ratio between conductance-derived stroke volume and a reference stroke volume, was used to relate both stroke volumes. The reason for this factor
was attributed to a non-homogeneous electrical field within the ventricle [2,8].
In various studies, parallel conductance was indeed found to depend on left ventricular volume [7,8,10] whereas in other studies no such dependency was found [15,18,24,25]. Recently, we described a new method to estimate parallel conductance (Gap), based on the determination of the area under a conductance dilution curve and the determination of flow and conductivity of blood, which appeared to be (four times) more precise than the conventional method [26]. In the present study, we extended the new conductance-dilution method to test the hypothesis that parallel conductance depends on left ventricular volume throughout the cardiac cycle. The non-homogeneous field has been shown to be the causal factor for the dependency of parallel conductance on left ventricular volume [23], as well as for the need for
[2]. Therefore, we also tested whether the hypothesis implies that a correction with
is not needed if parallel conductance is determined throughout the cardiac cycle.
First, we compared the estimates of real parallel conductances Gap and Glp, averaged over the cardiac cycles within the dilution curves. Secondly, Gap was determined throughout the cardiac cycle and its volume dependency was assessed. Thirdly, the factor
was calculated as the ratio between stroke volume, obtained by the conductance method using the conventional method to estimate parallel conductance, and that obtained by a thermodilution method.
| 2 Methods |
|---|
|
|
|---|
2.1 The conductance method
The general principles of the conductance method [2,6] and the configuration used for this study (7F conductance catheter (Sentron, The Netherlands), Leycom CFL512 (CD Leycom, Zoetermeer, The Netherlands)) have been described extensively elsewhere [26]. Also, the conventional method and the new approach to determine parallel conductance averaged over a cardiac cycle have been described elsewhere [2,6,26]. For this new approach, the indicator dilution method for hypertonic saline is described by a mass balance, in which the amount of ions injected is equal to the change in amount of ions detected. The amount of injected ions (indicator) influencing conductance is the product of the injected volume and the NaCl ion-concentration difference between the hypertonic saline and the blood before injection. The amount of detected ions is the product of flow and the integrated area under the NaCl ion-concentration, i.e. conductivity of blood mixed with NaCl, dilution curve. If the injection is given as a bolus and if blood flow (
|
| (1) |
i the electrical conductivity of the injectate at blood temperature,
b the conductivity of the blood before injection, 
b the change in conductivity of the blood due to the injection of the hypertonic saline. For left ventricular measurements, the volume of the left ventricle (Qv) between two measuring electrodes at any time t can be calculated as:
|
| (2) |
is assumed to be 1, Eqs. (1) and (2)
![]() | (3) |
i) and that of blood (
b) was determined using a measuring cell at body temperature. The experimental protocol of the measurements has been described elsewhere [26].
|
2.2 Subject population
The data were used from the same group of patients as described in a previous study [26]. We performed the study in 12 cardiac failure patients with dilated cardiomyopathy who had undergone cardiomyoplasty. Cardiomyoplasty gives a 40% reduction in left ventricular end-diastolic volume [19]. We obtained data in duplicate from all these 12 patients during left ventricular catheterization after cardiomyoplasty. Only measurements were used if the cardiac rhythm of the patient was regular during the measurement. The procedures performed in this study had been approved by the medical ethics committees of the various hospitals where the patients were accommodated [19]. The investigation conformed to the principles outlined in the Declaration of Helsinki [31]. Informed patient consent was obtained for insertion of catheters and subsequent measurements. The patients were heparinized before catheterization. A Swan Ganz thermodilution catheter was positioned in the pulmonary artery and a lumen of this catheter, positioned near the entrance of the right atrium, was used for the injection of hypertonic saline. The conductance catheter was inserted via a femoral artery into the left ventricle.
2.3 Data analysis
First, both estimates of parallel conductances Glp and Gap, averaged over the cardiac cycles within the dilution curve, were compared using a paired t-test. Secondly, Gap was determined for each of the 10 intervals within a cardiac cycle and plotted versus the corresponding left ventricular volume (QaV) and analysed using linear regression. To obtain stroke volume, end-diastolic minus peak systolic ventricular volume was calculated with the conductance method, assuming a constant parallel conductance Glp, and also with the time-varying parallel conductance Gap. Because left ventricular volume QaV is an average value of a certain interval within the cardiac cycle, the maximum and minimum QaV within the cardiac cycle will be slightly under- and over-estimated, respectively. Therefore, we used the maximum and minimum conductance found in each cardiac cycle during the dilution curve to calculate stroke volume QaV. This stroke volume was compared to stroke volume obtained by thermodilution (QTDV) using a paired t-test.
| 3 Results |
|---|
|
|
|---|
Baseline haemodynamic conditions such as heart rate and cardiac output as determined with thermodilution are shown in Table 1. In total, 24 measurements, i.e. 12 pairs, obtained in the left ventricle of 12 patients were used to calculate parallel conductance values with the conventional Glp-method and the new Gap-method. The values obtained by both methods to estimate parallel conductance averaged over various cardiac cycles were not significantly different, whereas the mean difference=4.04±6.3 S.E. (S).
|
Fig. 2 shows a typical example of the values of parallel conductance (Gap) and left ventricular volume (QaV) both plotted as a function of the 10 intervals occurring within the cardiac cycle. The first interval started at the end of diastole and the last interval ended at the end of systole. At the interval where left ventricular volume was maximal (usually interval 3, 4 or 5), parallel conductance was minimal for each measurement. The Gap averaged over a certain interval minus the Gap averaged over the whole cardiac cycle is calculated as percentages of the Gap averaged over the whole cardiac cycle. These data were averaged for all measurements (n=24) and plotted versus the corresponding intervals (Fig. 3). The error bars present the 95% confidence intervals. At end-systole (interval 1), the deviation from mean Gap was 6.59% and at peak-systole (interval 5) the deviation from mean Gap was –12.98%.
|
|
Using the same individual patient data as plotted in Fig. 2 as well as the data for the corresponding duplicate measurement, parallel conductance (Gap) was plotted versus left ventricular volume (QaV) (Fig. 4). The two different symbols indicate the two paired measurements. A linear regression line was the best fit through the values (Fig. 4). Parallel conductance (Gap) decreased with increasing left ventricular volume (QaV) (Fig. 4). No significant difference (paired t-test, P
0.05) was found between slopes and intercepts of duplicate measurements (n=12). The slopes of the 12 linear regression lines and their corresponding 95% confidence intervals of their estimates are given in Table 1.
|
3.1 Stroke volume and 1/

Fig. 5 clearly shows that stroke volume obtained from the new conductance-dilution method, assuming conductance to be volume-dependent, resembles stroke volume obtained with the thermodilution method better than stroke volume obtained with the conventional conductance method. Stroke volume obtained from the new conductance-dilution method (QaSV), did not differ significantly from stroke volume obtained from the thermodilution (QTDSV) (paired t-test, QTDSV–QaSV=1.90, 95% CI=–3.99, 7.80, S.E.=2.88 (ml)) (Fig. 5). Thus, left ventricular volume did not need a correction with 1/
(Eq. (2)), i.e.
is not significantly different from 1, if the new method to estimate parallel conductance is used. Stroke volume obtained by the conventional conductance method (QlSV) was significantly smaller than QTDSV (paired t-test, QTDSV–QlSV=46.07, 95% CI=35.89, 56.25 [ml]) (Fig. 5). Therefore, 1/
(QTDSV/QlSV) differed significantly from 1 (paired t-test, 1/
=3.09, 95% CI=2.40, 3.78).
|
| 4 Discussion |
|---|
|
|
|---|
Using an extension of a recently developed method to derive parallel conductance [26], we have shown in heart failure patients that parallel conductance decreased linearly within a cardiac cycle if left ventricular volume increased. In each patient, for a duplicate measurement, this relation between parallel conductance and left ventricular volume was similar, but between patients, variation in this relation was found (Table 1). This variation is probably due to variance in hydration level and therefore the conductivity of the tissues and the fluids surrounding the left ventricular cavity and other patient-related factors such as anatomy.
Parallel conductance was shown to be inversely related to ventricular volume (Fig. 3A), following the same pattern as that observed by others [15,24,25]. We found Gp to deviate significantly on average 13% of the mean at mid-systole, i.e. maximal volume. Szwarc and colleagues also found a significant decrease in Gp at mid-systole in the left ventricles of minipigs but this decrease was rather small [15]. No significant variation in parallel conductance within the cardiac cycle could be detected in the left ventricle of dogs [24] and in the left ventricle of children with a congenital cardiac disease [25].
In contrast to our results where only spontaneous cardiac variations were considered, it was found that if volume was varied artificially over a broad range then parallel conductance decreased with left ventricular volume [7,8]. Boltwood et al. reported that parallel conductance remained constant, if measured under steady-state conditions or measured during aortic or pulmonary arterial occlusion, but decreased with decreasing left ventricular volume during vena caval occlusion probably due to a decrease in the filling of the right ventricle [7]. Applegate et al. compared conductance of left ventricular volumes to those estimated from three orthogonal pairs of endocardially placed sonomicrometer crystals in open chest dogs [8]. This relation was reasonably linear during the volume excursions of a normal cardiac cycle, but non-linear if left ventricular volume was reduced below steady state end-systolic volume, implying that parallel conductance decreased with volume. Cassidy and Teitel [10] found a better correlation between total conductance volume, i.e. volume not corrected for parallel conductance, and volume obtained by cineangiography, than between ventricular blood volume, i.e. volume corrected for one constant volume-independent value of parallel conductance, and that obtained by cineangiography. They concluded that an error is introduced in the calculation of blood volume, if parallel conductance is assumed to be volume-independent.
The method to estimate parallel conductance within a cardiac cycle, which was used by the other groups [15,24,25] was derived from the Glp-method as described by Baan et al. [1]. Furthermore, to estimate average parallel conductance, the Glp-method [7,8,10] or a method derived from the Glp-method [14] was also used. We found previously that our method, based on the estimation of an area under a dilution curve and knowledge of flow, estimates parallel conductance averaged over various cardiac cycles with a factor four times better repeatability than the Glp-method [26]. We therefore assume that our method to determine parallel conductance within a cardiac cycle also estimates parallel conductance more precisely than the method derived from the Glp-method and that if a volume-dependency is present, this would be more easily detected with our method. Theoretical and experimental analysis of the area under the dilution curve revealed errors less than 1% made by neglecting the pulsatile character of blood flow [32]. Although this error contributes to the error made in the determination of parallel conductance within the cardiac cycle, we consider this error of minor importance. It should be noted that we determined left ventricular volumes of adult humans after cardiomyoplasty, whereas other studies were performed in the smaller left ventricles of children, dogs, or piglets [7,8,10,14,15,24,25], which might influence the slope of the relation between parallel conductance and volume. However, no relation was found between cardiac output (Table 1) and the slope of the relation between parallel conductance and left ventricular volume (Table 1). To correct for a possible influence of saline accumulation in the ventricular wall, passage of saline in the adjacent ventricle and recirculation on our calculated value of parallel conductance, we subtracted a baseline from the dilution curve. To correct for the influence of breathing, the dilution curve was recorded during a prolonged expiratory pause [26].
The correction factor
is assumed to be a correction for the inhomogeneity of the electrical field [1,8]. A non-homogeneous electrical field was suggested to cause parallel conductance to be volume-dependent [23]. Indeed, we found that if volume dependency of parallel conductance was taken into consideration then the correction with
was not needed because it was not significantly different from one.
To consider volume dependency, only slight changes in the conventional protocol for the determination of parallel conductance are necessary. Those include that the concentration and volume of the injectate should be known and that the injection and recording of the dilution curve should take place during an expiratory pause to eliminate the influence of breathing on the obtained conductivity dilution curve.
In conclusion, parallel conductance should be determined at each average volume level and throughout each cardiac cycle or the conductance of the left ventricle should be corrected with the factor
at the corresponding volume to eliminate a possible influence of electrical field heterogeneity.
Time for primary review 35 days.
| References |
|---|
|
|
|---|
- Baan J., Jong T.T.A., Kerkhof P.L.M., et al. Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter. Cardiovasc Res (1981) 15(6):328–334.
[Abstract/Free Full Text] - Baan J., van der Velde E.T., de Bruin H.G., et al. Continuous measurement of left ventricular volume in animals and humans by conductance catheter. Circulation (1984) 70(5):812–823.
[Abstract/Free Full Text] - Burkhoff D., van der Velde E., Kass D., et al. Accuracy of volume measurement by conductance catheter in isolated, ejecting canine hearts. Circulation (1985) 72(2):440–447.
[Abstract/Free Full Text] - Kass D.A., Yamazaki T., Burkhoff D., Maughan W.L., Sagawa K. Determination of left ventricular end-systolic pressure–volume relationships by the conductance (volume) catheter technique. Circulation (1986) 73(3):586–595.
[Abstract/Free Full Text] - Baan J., van der Velde E.T. Sensitivity of left ventricular end-systolic pressure–volume relation to type of loading intervention in dogs. Circ Res (1988) 62(6):1247–1258.
[Abstract/Free Full Text] - Baan J., van der Velde E.T., Steendijk P., Koops J. Calibration and application of the conductance catheter for ventricular volume measurement. Automed (1989) 11:357–365.
- Boltwood C.M.J., Appleyard R.F., Glantz S.A. Left ventricular volume measurement by conductance catheter in intact dogs. Parallel conductance volume depends on left ventricular size. Circulation (1989) 80(5):1360–1377.
[Abstract/Free Full Text] - Applegate R.J., Cheng C.P., Little W.C. Simultaneous conductance catheter and dimension assessment of left ventricle volume in the intact animal [see comments]. Circulation (1990) 81(2):638–648.
[Abstract/Free Full Text] - Teitel D.F., Klautz R., Steendijk P., et al. The end-systolic pressure–volume relationship in the newborn lamb: effects of loading and inotropic interventions. Pediatr Res (1991) 29(5):473–482.[Web of Science][Medline]
- Cassidy S.C., Teitel D.F. The conductance volume catheter technique for measurement of left ventricular volume in young piglets. Pediatr Res (1992) 31(1):85–90.[Web of Science][Medline]
- Odake M., Takeuchi M., Takaoka M., et al. Determination of left ventricular volume using a conductance catheter in the diseased human heart. Eur Heart J (1992) 13(Suppl_E):22–27.
[Abstract/Free Full Text] - Van der Velde E.T., van Dijk A.D., Steendijk P., et al. Left ventricular segmental volume by conductance catheter and Cine-CT. Eur Heart J (1992) 13(Suppl E):15–21.
[Abstract/Free Full Text] - Solda P.L., Pantaleo P., Perlini S., et al. Continuous monitoring of right ventricular volume changes using a conductance catheter in the rabbit. J Appl Phys (1992) 73(5):1770–1775.
[Abstract/Free Full Text] - Szwarc R.S., Mickleborough L.L., Mizuno S., et al. Conductance catheter measurements of left ventricular volume in the intact dog: parallel conductance is independent of left ventricular size. Cardiovasc Res (1994) 28(2):252–258.
[Abstract/Free Full Text] - Szwarc R.S., Laurent D., Allegrini P.R., Ball H.A. Conductance catheter measurement of left ventricular volume: evidence for nonlinearity within cardiac cycle. Am J Phys (1995) 268(4 Pt 2):H1490–H1498.
- Dickstein M.L., Yano O., Spotnitz H.M., Burkhoff D. Assessment of right ventricular contractile state with the conductance catheter technique in the pig. Cardiovasc Res (1995) 29:820–826.
[Abstract/Free Full Text] - White P.A., Bishop A.J., Conroy B., Oldershaw P.J., Redington A.N. The determination of volume of right ventricular casts using a conductance catheter. Eur Heart J (1995) 16:1425–1429.
[Abstract/Free Full Text] - White P.A., Chaturvedi R.R., Bishop A.J., et al. Does parallel conductance vary during systole in the human right ventricle. Cardiovasc Res (1996) 32:901–908.
[Abstract/Free Full Text] - Schreuder J.J., van der Veen F.H., van der Velde E.T., et al. Left ventricular pressure–volume relationships before and after cardiomyoplasty in patients with heart failure. Circulation (1997) 96(9):2978–2986.
[Abstract/Free Full Text] - Bishop A., White P., Oldershaw P., et al. Clinical application of the conductance catheter technique in the adult human right ventricle. Int J Cardiol (1997) 58(3):211–221.[CrossRef][Web of Science][Medline]
- Spinelli J.C., Valentinuzzi M.E. Conductivity and geometrical factors affecting volume measurements with an impedancimetric catheter. Med Biol Eng Comput (1986) 24:460–464.[CrossRef][Web of Science][Medline]
- Wu C.W., Skalak T.C., Schwenk T.R., et al. Accuracy of the conductance catheter for measurement of ventricular volumes seen clinically: effects of electric field homogeneity and parallel conductance. IEEE Trans Biomed Eng (1997) 44:266–277.[CrossRef][Web of Science][Medline]
- Kun S., Peura R.A. Analysis of conductance volumetric measurement error sources. Med Biol Eng Comput (1994) 32(1):94–100.[CrossRef][Web of Science][Medline]
- Lankford E.B., Kass D.A., Maughan W.L., Shoukas A.A. Does volume catheter parallel conductance vary during a cardiac cycle? Am J Phys (1990) 258(6 Pt 2):H1933–H1942.
- White P.A., Chaturvedi R.R., Shore D., et al. Left ventricular parallel conductance during cardiac cycle in children with congenital heart disease. Am J Phys (1997) 273:H295–H302.
- Kornet L., Schreuder J.J., van der Velde E.T., Baan J., Jansen J.R.C. A new approach to estimate parallel conductance for left ventricular volume measurements. Cardiovasc Res (2000) 48:455–463.
[Abstract/Free Full Text] - Jansen J.R.C., Schreuder J.J., Boogaard J.M., van Rooyen W., Versprille A. Thermodilution technique for measurement of cardiac output during artificial ventilation. J Appl Phys (1981) 50(3):584–591.
- Jansen J.R.C., Boogaard J.M., Versprille A. Extrapolation of thermodilution curves obtained during a pause in artificial ventilation. J Appl Phys (1987) 63(4):1551–1557.
[Abstract/Free Full Text] - Jansen J.R.C., Hoorn E., van Goudoever J., Versprille A. A computerized respiratory system including test functions of lung and circulation. J Appl Phys (1989) 67(4):1687–1691.
[Abstract/Free Full Text] - Jansen J.R.C., Schreuder J.J., Settels J.J., Kloek J.J., Versprille A. An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med (1990) 16(7):422–425.[CrossRef][Web of Science][Medline]
- World Medical Association. World Medical Association Declaration, Helsinki. Recommendations guiding physicians in biomedical research involving human subjects. Cardiovasc Res (1997) 35(1):2–3.
[Free Full Text] - Scheuer-Leeser M., Morguet A., Reul H., Irnich W. Some aspects to the pulsation error in blood-flow calculations by indicator-dilution techniques. Med Biol Eng Comput (1977) 15(2):118–123.[CrossRef][Web of Science][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||



) and parallel conductance (Gap) (
) are plotted versus the corresponding 10 equally sub-divided time-intervals within the cardiac cycle.